Pertactin:
Molar
abs. coefficient = (# trp) (5,500) + (# tyr) (1490) + (# cys) (125)
Trp = 9
Tyr = 7
Cys = 0
Molar abs. coefficient
= 59,930 M-1 cm-1
C = A / (E L) C = conc. of
protein (M) A = abs at 280nm E = molar abs. coeff.
L = path length (cm)
C = A / 59,930
MW = 61,400 g / mol
C = 1.02 A = mg / ml
Used 40 ug / ml of Pertactin for samples (this corresponded to 50 ul of an ion exchange fraction from Prep #1).
Volume: 900 1000 ul in cuvette (semi-micro).
Procedure:
1. Sign in to Fluorometer LogIn sheet located next to fluorometer. Follow
instruction on sheet.
2. Lamp needs to be warmed up for 20-30 min prior to use.
3. Make sure everything is off including computer and monitor.
4. Flip switch to turn lamp on. Then push ignite button. Note that the light
turns on and the fan is running. Allow time for warm-up.
5. Set slit widths to appropriate values.
6. Pull open red knob door located on the end of the monochrometer near the
lamp.
7. Turn on computer and monitor.
8. While booting up
.
9. Turn on Photomultiplier tube cooling water.
10. Turn on water alarm.
11. Turn on cuvette bath.
12. Turn on instrument.
13. Back at computer press any key until the option menu appears. Choose A for
OS/2. Then type y and press enter.
14. Close all windows that open at start.
15. Double click SLM 8100 icon in top left corner of screen. Program loads and
instrument initializes.
16. Click user tab, login, choose user id.
17. File Tab : ASCII export setup. Click ASCII extension is txt.
18. Status: Leave this window open while using instrument.
19. Channels : EmL is already loaded, add EmL / Ref, and add Ref. OK.
20. Monos : Excitation 280, EmL 330.
21. Shutters : Excitation=Open, Check=Open Shutter for Acquisition, Emission
Left=Open, Emission Right=Closed, Reference=Open.
22. Sensitivity : Ref click AR at Gain of 1 this will automatically set the
reference. Put in a blank sample and click AR this will automatically set the
EmL. May need to change the Gain from 1 to 10 or 100. Record Gains and HV for
each.
23. Go to Applications tab : Emission scan : set lower and upper limits 300
and 400. Set integration, scan rate, # of repetitions. OK.
24. Go to Applications tab : Start Application : Select File name and type.
25. When finished go to File : Open. Choose search and change from *.dat to
*.txt. Select files and click Copy. Type a: to copy to diskette.
26. Make sure all data is saved.
27. Exit SLM 8100 program and double click Dual Boot. Type y and enter. Computer
will begin shutdown process. Meanwhile
28. Follow fluorometer sheet for shutting things down in reverse order.
29. Then turn off computer and very last turn off the lamp and record the total
time used.